Section 1 —Introduction
Page 1-4
VitalCare
TM
506N3 Series Service Manual
Criticare Systems, Inc.
Cuff Inflation and
Pressure Protection
The maximum allowable cuff pressure is 300 mmHg adult mode (150
mmHg in neonate mode). Adult mode cuff pressure is allowed to
remain above 15 mmHg for a maximum of 180 seconds. Neonatal
mode cuff pressure is allowed to remain above 5 mmHg for a
maximum of 90 seconds.
The monitor automatically deflates the cuff if the time limit is violated.
The monitor contains hardware protection for overpressure
conditions, pressure transducer failures, or microprocessor and pump
control circuit failures.
Figure 1-1: NIBP Cuff Pressure and Pulse over Time
B.P Cuff Inflation Pressure
(Shown during inflation)
Systolic Pressure
Cuff deflates rapidly after
monitor determines systolic
pressure
Time
Diastolic
Pressure
Mean Pressure
Systolic
Pressure
Pulse Waveform
(Measured from B.P. Cuff
Pressure Fluctuation)
Pressure in
mmHg
Diastolic Pressure
Actual Blood
Pressure Waveform
Summary of Contents for 506DN3
Page 12: ......
Page 56: ......
Page 62: ......
Page 90: ......
Page 156: ......
Page 184: ......
Page 225: ......
Page 229: ......
Page 230: ......
Page 231: ......
Page 232: ......
Page 233: ......
Page 237: ......
Page 238: ......
Page 239: ......
Page 240: ......
Page 241: ......
Page 242: ......
Page 243: ......
Page 244: ......
Page 245: ......
Page 247: ......
Page 248: ......
Page 249: ......
Page 250: ......
Page 251: ...DO NOT SCALE PRINT DO NOT SCALE PRINT REV DATE DESCRIPTION BY...
Page 253: ......
Page 254: ......