Service Manual
Page 7-27
Section 7 —Disassembly
11.Fold over the membrane’s ground tail and connect to the Main
Board using a P.H.M.S. screw and washer. Attach the right side
of the bezel and torque all five (5) screws at 5 in. lbs.
Figure 7-32: Membrane Ground Tail
If the monitor previously contained a pn 91384A001 or
91384A002 Main Board that was replaced with a 91384A003
Main Board:
a. Reuse the black ground wire that was removed in Step 3
by inserting P.H.M.S. screws through the ring terminal
(lower right corner) and both holes of the ground tail. (This
is done in place of the washer stated above.)
b. Remove the green ground wire from the metal chassis.
Figure 7-33: Remove Ground Wire from Chassis
(Main Board Upgrades)
If the monitor contained a pn 91384A001 or 91384A002 Main
Board and the replacement is a pn 91384A001 or 91384A002
Main Board:
a. Reuse the black ground wire that was removed in Step 3
by inserting P.H.M.S. screws through the ring terminal
(lower right corner) and both holes of the ground tail. (This
is done in place of the washer stated above.)
b. Reconnect the green ground wire at
P3
on the Main Board.
Membrane
Ground Tail
Remove green
ground wire
Summary of Contents for 506DN3
Page 12: ......
Page 56: ......
Page 62: ......
Page 90: ......
Page 156: ......
Page 184: ......
Page 225: ......
Page 229: ......
Page 230: ......
Page 231: ......
Page 232: ......
Page 233: ......
Page 237: ......
Page 238: ......
Page 239: ......
Page 240: ......
Page 241: ......
Page 242: ......
Page 243: ......
Page 244: ......
Page 245: ......
Page 247: ......
Page 248: ......
Page 249: ......
Page 250: ......
Page 251: ...DO NOT SCALE PRINT DO NOT SCALE PRINT REV DATE DESCRIPTION BY...
Page 253: ......
Page 254: ......