Criticare Systems, Inc.
VitalCare
TM
506N3 Series
Service Manual
Page 7-41
Section 7 —Disassembly
Replace ComfortCuff
NIBP Module
First perform “Chassis Service/Pump Disassembly” procedure.
1. With the #2 Phillips screwdriver, remove the four (4) screws
(pn 40496B004) holding the chassis assembly in place.
Figure 7-53: Remove Four Screws (Chassis)
2. Slide the chassis out of the rear housing and pull the NIBP
fitting from the opening on the right side of the rear housing.
Keep temp and SpO
2
cables (temp units) to the right.
3. Remove the ground wire from
JP1
.
4. Remove the harness from
CN1
.
5. Flip the chassis up with the power supply facing up.
6. Loosen the four (4) screws (pn 40496B001) that are contained
with the four (4) #6 cup washers (pn 42499B001). Lift the power
supply PCB (pn 86517C001) up and remove the four cup
washers that lie beneath the power supply.
7. Loosen the four (4) screws (pn 40995B006) attached to the
chassis beneath the power supply. Torque at 5 in. lbs.
8. Lift the NIBP module up. (The direction of the fitting faces the
same direction as the shape of the chassis opening.
9. Unplug the NIBP cable (pn 90926A002) from the NIBP module
(pn 93947A003) at
J1
. Note that the connector is locking.
10.Replace the NIBP module.
11.Reassemble in reverse order.
Reassemble monitor and follow “Completion of Service” procedure at
the end of this section.
Remove Screws
Summary of Contents for 506DN3
Page 12: ......
Page 56: ......
Page 62: ......
Page 90: ......
Page 156: ......
Page 184: ......
Page 225: ......
Page 229: ......
Page 230: ......
Page 231: ......
Page 232: ......
Page 233: ......
Page 237: ......
Page 238: ......
Page 239: ......
Page 240: ......
Page 241: ......
Page 242: ......
Page 243: ......
Page 244: ......
Page 245: ......
Page 247: ......
Page 248: ......
Page 249: ......
Page 250: ......
Page 251: ...DO NOT SCALE PRINT DO NOT SCALE PRINT REV DATE DESCRIPTION BY...
Page 253: ......
Page 254: ......