Section 6 —Service Testing & Calibration
Page 6-8
VitalCare
TM
506N3 Series
Service Manual
Criticare Systems, Inc.
Calibrate
1. Select
Calibrate
from the service tool screen on the PC.
Figure 6-6: Select Calibrate
2. A box appears as shown. Verify that the pressure at the cuff is
“0mmHg” ±2mmHg. If not, adjust R1 for 0.122 @ TP 10.
NOTE
: Adjustment is only necessary for older NIBP Boards
(pn 91325A002). NIBP Boards (pn 91325A003 and later) are
self-calibrated to “0.”
.
Figure 6-7: Verify Cuff Pressure