Criticare Systems, Inc.
VitalCare
TM
506N3 Series
Service Manual
Page 7-31
Section 7 —Disassembly
Replace LCD and LED
Display Boards
First perform “Disassemble Front Bezel from Rear Housing”
procedure.
1. Remove the speaker from the rail of the front bezel. Lift the
orange and brown wires from the relief of main as shown below.
Figure 7-38: Remove Speaker
2. Unplug the speaker assembly (pn 90932A001) from
J3
of the
Main Board. Place speaker assembly aside.
3. With a #1 Phillips screwdriver, remove the screw
(pn 40995B005) and washer (40086B002) holding the folder
tab from the membrane to the Main Board.
Figure 7-39: Membrane Tab
NOTE
: If the monitor has an older Main Board (pn 91384A001
or 91384A002), remove the two (2) screws holding the ground
wire to the Main Board. Disconnect the green ground wire from
P3
on the Main Board.
Speaker
Folded Tab
from Membrane