Section 5 — Preventative Maintenance
Incoming Inspection
You should inspect all new monitors upon arrival for shipping damage
before you place them into operation. The monitor should be free
from dents, cracks, or other physical damage. The quality inspection
seal of the monitor should be unbroken; this indicates that the
monitor has been tested according to the manufacturer’s
specifications.
If further incoming inspection or testing is desired, the manufacturer
recommends you use “Speaker Performance, Alarms Verification” in
this section as an incoming inspection test. You may perform
additional electrical safety testing with “Electrical Safety Tests” in this
section as part of an incoming inspection in accordance with the
policies of the health care provider.
Maintenance Schedule
Long-Term Storage
No special preparation is necessary for long term storage of the
monitor. Although the battery does not have to be removed from the
monitor for long term storage, the battery does drain to an
unrecoverable state after 3 months without periodic charging.
Service Checks
If the monitor shows any signs of physical damage, contact the
Criticare Service Department for repair.
Technical Service (US): (800) 458-2697
International Customer Service: (262) 798-8282
Calibration
No periodic calibration of the monitor is necessary. It is
recommended to perform an NIBP calibration verification as part of
the annual safety testing.
Every Patient
• Clean and disinfect the NIBP cuff and the SpO
2
cable as
needed.
• Inspect the accessories for damage.
Every Day
• Charge the monitor’s battery as necessary.
Every 3 Months
• Clean the exterior of the unit (or clean as needed).
• Inspect the monitor and AC (mains) cord for damage.
Every Year
• Perform the annual safety tests that are described in this
section.
Summary of Contents for 506DN3
Page 12: ......
Page 56: ......
Page 62: ......
Page 90: ......
Page 156: ......
Page 184: ......
Page 225: ......
Page 229: ......
Page 230: ......
Page 231: ......
Page 232: ......
Page 233: ......
Page 237: ......
Page 238: ......
Page 239: ......
Page 240: ......
Page 241: ......
Page 242: ......
Page 243: ......
Page 244: ......
Page 245: ......
Page 247: ......
Page 248: ......
Page 249: ......
Page 250: ......
Page 251: ...DO NOT SCALE PRINT DO NOT SCALE PRINT REV DATE DESCRIPTION BY...
Page 253: ......
Page 254: ......