Section 2 —Service Menus
Page 2-2
VitalCare
TM
506N3 Series
Service Manual
Criticare Systems, Inc.
Secondary Service
Displays
Pressing the MENU key once within a primary service menu item
accesses the secondary service menu item(s).
With the exception of the
REVISIONS
service item, an arrow (<--)
points at the secondary service menu item to designate that the user
is in this mode. Press UP and DOWN arrow keys to scroll through
secondary service menu items.
Press the MENU key to choose the desired secondary service menu.
This toggles the arrow to point the other way {-->} and moves to the
value selection.
Hold the MENU key to return to the primary service menu item.
Service Display List
The following is a list of service primary and secondary menu items
follows and each default and description/values.
Primary Menu
Secondary Menu
Default
Description
REVISIONS
506N3_SERIES
N/A
Model series
Revision
nn.nn
N/A
Main software revision
nn.nn
is alphanumeric
revision identifier – decimal point may vary
MMM DD YYYY
N/A
Main software revision date
S/N:
nnnnnnnnn
N/A
Monitor serial number
APP. CS. :
nnnn
N/A
Main software checksum
SpO
2
XXXXX
nn.nn
N/A
SpO
2
module revision where
XXXXX
is a three to
five character name for the type of SpO
2
module;
nn.nn
is alphanumeric revision identifier – decimal
point may vary.
SpO
2
NOT SENT
N/A
SpO
2
module did not send software revision
information. This message displays if no module
is present.
NIBP
XXXXX
nn.nn
N/A
NIBP module revision where
XXXXX
is a three to
five character name for the type of NIBP module;
nn.nn
is alphanumeric revision identifier – decimal
point may vary.
NIBP NOT SENT
N/A
NIBP module did not send software revision
information. This message displays if no module
is present.
TEMP
XXXXX nn.nn
N/A
Temperature module revision where
XXXXX
is a
three to five character name of the type of
temperature module;
nn.nn
is alphanumeric
revision identifier – decimal point may vary.
TEMP NOT SENT
N/A
Temperature module did not send software
revision information. This message displays if no
module is present.
NOTE
: When entering Revisions, if accessed to quickly some
revisions may not show. Wait 5-10 seconds before entering to allow
the processor to acknowledge the revisions. If a revision does not
show, either that module is not present in the device or the processor
is not functioning correctly.
Summary of Contents for 506DN3
Page 12: ......
Page 56: ......
Page 62: ......
Page 90: ......
Page 156: ......
Page 184: ......
Page 225: ......
Page 229: ......
Page 230: ......
Page 231: ......
Page 232: ......
Page 233: ......
Page 237: ......
Page 238: ......
Page 239: ......
Page 240: ......
Page 241: ......
Page 242: ......
Page 243: ......
Page 244: ......
Page 245: ......
Page 247: ......
Page 248: ......
Page 249: ......
Page 250: ......
Page 251: ...DO NOT SCALE PRINT DO NOT SCALE PRINT REV DATE DESCRIPTION BY...
Page 253: ......
Page 254: ......