Section 5 —Preventative Maintenance
Page 5-26
VitalCare
TM
506N3 Series
Service Manual
Criticare Systems, Inc.
Functional and Safety Testing Checklist (Page 2 of 3)
PASS
FAIL
N/A
Functional Tests
Alarm Functions
Verify Pulse volume operation from 0 to 10
_____
_____
Verify Alarm volume operation from 0 to 10
_____
_____
Verify 2 minute Alarm Silence function
_____
_____
Verify permanent Alarm Silence function
_____
_____
Display Functions
Verify Green Charge LED w/AC power connected
_____
_____
Powerup Function
Verify unit powers up with AC plug in
_____
_____
Verify unit powers up on battery
_____
_____
NIBP Function
Simulator set to Adult mode @120/80 – 1 min cycle
(Each reading does not vary by more than ±4% or ±4mmHg (whichever is greater) from the calculated
average.)
_____/____, _____/____, _____/____, _____/____
Average: _____/____
_____
_____
Simulator set to Neonate mode @80/50 – Stat mode
(Each reading does not vary by more than ±4% or ±4mmHg (whichever is greater) from the calculated
average.)
_____/____, _____/____, _____/____, _____/____
Average: _____/____
_____
_____
Seal Test
_____
_____
SpO
2
Functions
Verify
SPO2: NO SENSOR
condition (DOX, NELLCOR)
_____
_____
Verify
SPO2: SENSOR
condition (DOX, NELLCOR)
_____
_____
Verify
SPO2: HIGH AMBIENT
condition (DOX only)
_____
_____
_____
Take SPO2 & HR readings (DOX, NELLCOR)
_____
_____
Verify LED Bargraph (DOX, NELLCOR)
_____
_____
Sensor symbol operation (DOX, NELLCOR)
_____
_____
Verify
SPO2: LOST
condition (DOX only)
_____
_____
_____
Verify
SPO2: PULSE LOST
condition (NELLCOR only)
_____
_____
_____
FasTemp Temperature Functions
_____
Verify
TEMP: NO PROBE
condition
_____
_____
Verify Temperature readings Blue box Oral
_____
_____
Verify Temperature readings Blue box Axial
_____
_____
Verify Temperature Simulator readings 37.0 C or 98.6 F
_____
_____
Verify Temperature Readings Red box Rectal
_____
_____
TurboTemp Temperature Functions
_____
Verify
TEMP: NO PROBE
appears
_____
_____
Connect the Alaris TE 1811 probe and LED blinking
_____
_____
Verify 98 degrees at the
Start
position
_____
_____
Verify that the temp reads 80.2 F ±0.2 or 26.8 C ±0.1
_____
_____
Verify that the temp reads 98.0 F ±0.2 or 36.7 C ±0.1
_____
_____
Verify that the temp reads 98.6 F ±0.2 or 37.0 C ±0.1
_____
_____
Verify that the temp reads 102.0 F ±0.2 or 38.9 C ±0.1
_____
_____
Verify that
TEMP: INVALID
appears @ 107.8
_____
_____
Verify that BP reads 106.0 degrees or ---.-F
_____
_____
Verify that
TEMP: BAD PROBE
appears when BP is pressed _____
_____
Verify that the rectal LED illuminates when probe is inserted
and withdrawn
_____
_____
Summary of Contents for 506DN3
Page 12: ......
Page 56: ......
Page 62: ......
Page 90: ......
Page 156: ......
Page 184: ......
Page 225: ......
Page 229: ......
Page 230: ......
Page 231: ......
Page 232: ......
Page 233: ......
Page 237: ......
Page 238: ......
Page 239: ......
Page 240: ......
Page 241: ......
Page 242: ......
Page 243: ......
Page 244: ......
Page 245: ......
Page 247: ......
Page 248: ......
Page 249: ......
Page 250: ......
Page 251: ...DO NOT SCALE PRINT DO NOT SCALE PRINT REV DATE DESCRIPTION BY...
Page 253: ......
Page 254: ......