Maintenance, Troubleshooting and Warranty
Troubleshooting
Rev. D
Dimension 3100 Manual
347
•
“
Illuminator Error, Manual override is on
” – The manual control on the front of the
D3100 stage controller is being used. Turn the illuminator manual control knob counter-
clockwise until a click sound is heard, indicating that it is switched off.
•
“
Bad data read from illuminator was 0,S.B. 80
” – The D3100 stage controller power
is off. The power supply inside the D3100 stage controller is not working.
•
“
Warning focus position not updated
” – You tried to move X-Y stage with no AFM
SUM signal.
•
“
Safety check failed SUM too low
” – 2 Explanations:
•
You tried to engage with no AFM SUM signal.
•
You tried to move X-Y stage with no SUM signal.
•
“
The focus axis is at its limit
” – The focusing objective is either fully extended or fully
retracted. This message will appear during the locate tip routine.
•
“
Warning: Sample clearance reduced! Unable to move past limit
” – The focusing
objective reached the extended limit when moving to the locate tip position. This can
happen during STM use due to the lower tip location. The end of the STM probe should
protrude no more than 1mm below the end of the tip holder tube. This message can also
occur if the sample clearance is set to a very large value. The standard sample clearance
distance is 1000µm.
•
“
Motor axis is at limit
” – The SPM (Z stage) is fully retracted. This can happen after
the stage is initialized. Ignore the message and lower the Z stage by using the Focus
Surface controls.
•
“
The tip location has been changed to an unsafe distance. If you proceed with this
change you will need to re-teach the tip replacement position.
” – This warning
appears during the Locate Tip routine if the focusing objective has moved more than
300µm from the previous tip focus position. This is common when switching between
silicon and silicon nitride cantilevers due the thickness difference of their substrates.
The message will also appear when switching between AFM and STM probes. The
intent of this message is a safety check to warn you that you have moved the objective
unusually long distance. Select
OK
if you’ve changed to a different style of probe.
Summary of Contents for Dimension 3100
Page 12: ...xii Dimension 3100 Manual Rev D ...
Page 20: ...List of Figures xx Dimension 3100 Manual Rev D ...
Page 72: ......
Page 106: ......
Page 118: ......
Page 214: ...Scanning Tunneling Microscopy STM Etching Tungsten Tips 194 Dimension 3100 Manual Rev D ...
Page 222: ......
Page 266: ......
Page 274: ......