Force Imaging
Force Calibration (Contact Mode AFM)
220
Dimension 3100 Manual
Rev. D
Motor
The
Motor
menu allows you to withdraw or manually control the tip position using the stepper
motor. Selecting
Step Motor
opens a dialog box containing the following buttons:
•
Tip Up
:
This command moves the tip up by the
SPM step size
displayed inside the
window.
•
Tip Down
:
This command moves the tip down by the
SPM step size
displayed inside
the window.
13.4 Force Calibration (Contact Mode AFM)
13.4.1 Obtaining a Good Force Curve
Figure 13.4a
Typical Force Calibration Curve
Figure 13.4b
Piezo Positions for Typical Force Curve
To minimize or calculate the contact force between the tip and sample, obtain a good force curve
which shows the typical features displayed in the example curve in
. However, the
force curve rarely looks “typical” right after invoking
Force Calibration
mode. This section
(Z Scan start -Ramp size)
Setpoint
Slope = Volts of Deflection/nanometers (or volts) of piezo travel
+
Cantilever Off Surface
(Z Scan start)
Z Piezo Voltage
1
2
3
4
5
6
V csmin
Cantile
v
er Deflection
V
oltage
Piezo travel
direction
Position
number
Up
1
Up
2
At
Bottom
3
Down
5
6
Down
4
Up
Up
Summary of Contents for Dimension 3100
Page 12: ...xii Dimension 3100 Manual Rev D ...
Page 20: ...List of Figures xx Dimension 3100 Manual Rev D ...
Page 72: ......
Page 106: ......
Page 118: ......
Page 214: ...Scanning Tunneling Microscopy STM Etching Tungsten Tips 194 Dimension 3100 Manual Rev D ...
Page 222: ......
Page 266: ......
Page 274: ......