TappingMode AFM
Basic TappingMode AFM Operation
138
Dimension 3100 Manual
Rev. D
Figure 9.2c
Cantilever Tune Frequency Sweep
11. Click
OK
. The parameters set in the
Cantilever Tune
control panel appear in the
Realtime
control panel.
12. Click on
CANCEL
to exit the
Cantilever Tune
command and leave the parameters
unchanged.
Cantilever
Response
0.50 V/div
Center Frequency - 337.8723 KHz
0.05 KHz/div
Setpoint
Frequency Sweep