Stage System
Stage Menu Commands
Rev. D
Dimension 3100 Manual
75
Figure 5.2d
Focus Surface Prompt
2. Use the left button on the trackball to focus on the surface (which moves the SPM or Z stage
up and down).
3. Use the right button on the trackball to zoom out completely when trying to focus on the
surface.
4. If the sample is not already under the microscope, use the trackball (with neither the left nor
right buttons pressed) to move the X-Y stage until the sample is in position.
5. If the surface is partially in focus, use the
Autofocus
option to complete the focusing
process.
6. To focus on the sample “surface” (normal operation) or the “tip reflection” (for extremely
clean samples), change the
Focus On
parameter accordingly.
Note:
For reflective or semi-reflective samples, the tip reflection is easier to bring into
focus than the surface, especially if the sample is very flat or clean.
7. For samples which are difficult to bring into focus, move to an edge of the sample, which is
easy to find in the optical image, and bring the top of the edge into focus.
8. Move the sample back to the desired X-Y position.
9. Verify that the surface remains in focus.
CAUTION:
When moving the SPM stage up and down, it is possible to crash
the tip into the surface. To prevent a crash while focusing on the
surface, watch the optical image
and
tip-to-sample proximity. The
sample should be in focus when the tip is 1mm (1000µm) above
the surface.
027
Summary of Contents for Dimension 3100
Page 12: ...xii Dimension 3100 Manual Rev D ...
Page 20: ...List of Figures xx Dimension 3100 Manual Rev D ...
Page 72: ......
Page 106: ......
Page 118: ......
Page 214: ...Scanning Tunneling Microscopy STM Etching Tungsten Tips 194 Dimension 3100 Manual Rev D ...
Page 222: ......
Page 266: ......
Page 274: ......