Electric Techniques
Electric Force Microscopy
Rev. D
Dimension 3100 Manual
285
Jumper Configurations With Extender Electronics
As shipped from the factory, systems with the Extender Electronics option should have an original
backplane jumper configuration as shown in
. This configuration connects both the tip
and sample to the ground.
Figure 16.2j N
ormal Jumper Configuration with Extender Electronics Module
CAUTION:
Turn off the NanoScope controller and unplug the power cable
from the microscope electronic box before attempting to adjust
jumper configurations.
ChuckGround/Bias
Analog 2 or
Unused
Auxiliary 2 Output
Ground
STM Signal
Indicates Jumpers
Gnd/OSC + DC signal
Unused
Ground
Ground
AFM Tip
Chuck
Sample
Tip