4.9.3
LED – IrDA PWB
•
Remove the detector’s main cover.
•
The LED – IrDA PWB is located behind the front panel in
the top left hand corner as you face the detector.
•
Disconnect the ribbon cable to this PWB.
•
Locate and remove the two screws holding the PWB
and its mounting bracket. The PWB can now be
removed.
4.9.4
Power Supplies
•
Remove the main detector cover.
•
Remove the single screw holding down the power supply
module.
This is found on the outer edge of the detector adjacent
to the power supply module.