ProStar325 Service Manual
Publication date 8 May 2003
103
Replacement
The replacement procedure is the opposite of the removal
procedure.
•
Carefully place the lamp assembly into position
•
Tighten the two thumbscrews
•
Re-connect the three way connector
•
Engage the safety latch by sliding it back down.
5.3.2
Vis lamp source
Removal
•
Remove the front green door which will expose a flat
panel.
•
Remove this flat panel by loosening the holding screw on
the top left hand corner.
•
Locate the Vis lamp housing.
•
Loosen the thumbscrew and carefully remove the
housing.
thumbscrew