ProStar325 Service Manual
Publication date 8 May 2003
145
A small graph will appear indicating that mono calibration is in
progress. Once completed the “Mono Calibration completed” pop-
up box will appear. Press 'OK'.
The wavelength calibration is now complete.
Go back to 'Main' tab and un-tick “Enter Diagnostic State”.
7.12.2
0% Transmission
There are three 0%T calibration options available.
Calibrate with Vis lamp on
StockCheck.com