ProStar325 Service Manual
28
Publication date 8 May 2003
2.6
Electrical specifications
2.6.1
Mains supply
The installation of electrical power supplies must comply with
the rules and/or regulations imposed by local authorities
responsible for the supply of electrical energy to the workplace.
Warning
Good electrical grounding is essential to avoid potentially
serious shock hazards.
A 3-wire outlet with ground connection must be provided for
the detector.
Make certain that power outlets are earth-grounded at the
grounding pin.
All power supplies must be single-phase AC voltage, three-wire
system (active, neutral, earth) and should be terminated at an
appropriate power outlet receptacle that is within reach of the
detector power cord assembly. For safety reasons, a separate
power outlet receptacle should be provided for each unit in the
system. Do not plan to use extension cords or outlet adaptors.
All ProStar 325 detectors are supplied with a 2metre (6’ 6") long
power cord and three-pin plug assembly that is compatible with
common standards applicable in the local area.
Avoid using power supplies from a source that may be subject to
electrical or RF interference from other services (large electric
motors, elevators, and welders for example).
System Unit
Required Supply Voltage
Power
Rating
(typical
)
Detector
ProStar 325
100-240VAC ±10%,
50-60Hz +/-1Hz
130 VA
Computer
100, 120, 220, 240 VAC
50/60 Hz
300 VA
Printer
100, 120, 220, 240 ±10%
VAC 50/60 ±1 Hz
100 VA
Mains voltage requirements
Note
For non-Varian products, the table above is indicative only.
Refer to the literature provided with the computer, printer etc.
for details of individual power requirements.
StockCheck.com
Summary of Contents for ProStar 325 UV-Vis
Page 9: ...ProStar325 Service Manual Publication date 8 May 2003 9 S t o c k C h e c k c o m ...
Page 19: ...ProStar325 Service Manual Publication date 8 May 2003 19 S t o c k C h e c k c o m ...
Page 34: ...ProStar325 Service Manual 34 Publication date 8 May 2003 S t o c k C h e c k c o m ...
Page 71: ...ProStar325 Service Manual Publication date 8 May 2003 71 S t o c k C h e c k c o m ...
Page 72: ...ProStar325 Service Manual 72 Publication date 8 May 2003 S t o c k C h e c k c o m ...
Page 118: ...ProStar325 Service Manual 118 Publication date 8 May 2003 S t o c k C h e c k c o m ...
Page 149: ...ProStar325 Service Manual Publication date 8 May 2003 149 S t o c k C h e c k c o m ...