ProStar325 Service Manual
Publication date 8 May 2003
133
To activate these Analog tests you must click on Start Analysis.
Digital Toggles -
Verify the operation of the Sync signals
located at the rear of the detector. Toggle any signal by clicking
in its tick box and then monitor it at the output either on the D-
range connector pin, or on the Rear Panel Interface PWB if fitted.
The five detector relays can be individually toggled to verify their
operation as well as checking the bi-color status of each of the
three front panel LED’s.
7.10.2
EEPROM
Use this page to view and reset the information contents of the
EEPROM. From here you can reset the dates and operating
hours whenever installing a new lamp source, as well as the
installation and maintenance dates for the detector.
StockCheck.com
Summary of Contents for ProStar 325 UV-Vis
Page 9: ...ProStar325 Service Manual Publication date 8 May 2003 9 S t o c k C h e c k c o m ...
Page 19: ...ProStar325 Service Manual Publication date 8 May 2003 19 S t o c k C h e c k c o m ...
Page 34: ...ProStar325 Service Manual 34 Publication date 8 May 2003 S t o c k C h e c k c o m ...
Page 71: ...ProStar325 Service Manual Publication date 8 May 2003 71 S t o c k C h e c k c o m ...
Page 72: ...ProStar325 Service Manual 72 Publication date 8 May 2003 S t o c k C h e c k c o m ...
Page 118: ...ProStar325 Service Manual 118 Publication date 8 May 2003 S t o c k C h e c k c o m ...
Page 149: ...ProStar325 Service Manual Publication date 8 May 2003 149 S t o c k C h e c k c o m ...