5-21
SECTION 5
TEST RESULTS
STRATUM 2
Stratum 2 Clock Traceable
— A received
BPM stating timing is recognized as a Stratum
2 clock.
STRATUM 3
Stratum 3 Clock Traceable
— A received
BPM stating timing is recognized as a Stratum
3 clock.
STRATUM 4
Stratum 4 Clock Traceable
— A received
BPM stating timing is recognized as a Stratum
4 clock.
SYNC UNKNOWN
Synchronization Quality Unknown
— A re-
ceived BPM stating the synchronization timing
is either unknown or the signal is asynchronous.
USER
User BPM Detected
— A received BPM stat-
ing timing is recognized as an eight-bit
user-programmable BPM.
SLC ALM SEC
SLC Alarm Seconds
— This is a count of the
seconds during which one or more SLC-96
Datalink Alarms occurred. SLC ALM SEC is
displayed as a number from 0 to 99999999, and
it is updated once per test second. If the value
exceeds the displayed maximum, a greater than
sign (>) is inserted on the left and the displayed
digits remain accurate.