SECTION 2
2-16
INSTRUMENT CHECKOUT
AND MAINTENANCE
Smart Loopback/Command Codes Option
(107A-4)
If the Smart Loopback/Command Codes Op-
tion is installed, perform the following procedure to
verify its operation.
1.
Test setup
Prepare the T-BERD 107A for testing in
accordance with the Instrument Checkout
Basic Setup procedure.
2.
AUX and PATTERN switch
Press the
AUX
switch to activate the auxiliary
functions. Press the
PATTERN
switch to
verify the AUX SMARTNET function is
available.
3.
CATEGORY and RESULTS switches
Press the
CATEGORY
switch to select ILR
(intelligent line repeaters). Press the
RE-
SULTS
switch to select an intelligent line
repeater manufacturer and model (e.g.,
TELTREND 7239 LD). Refer to Section 4.2
Auxiliary Functions for additional informa-
tion on available intelligent repeaters.
AUX
SMARTNET
ILR
TELTREND
7239 LD
MODE
PATTERN
CATEGORY
RESULTS
MODE
CATG
PATT