Updating the Firmware
The PS Range Module contains two different computer boards:
•
The MPU board is responsible for the measurement process
•
The communication board is used for the Ethernet communication.
Both boards can be updated over the serial RS232 interfaces by use of the
XMODEM protocol. The sensor supports XMODEM-CRC and XMODEM-1K variants
to improve performance and transfer safety.
ATTENTION
ATTENTION
ATTENTION
ATTENTION
Carefully read this entire instructions before you start updating your sensor.
Do not power off the sensor until the firmware is updated!
8.1
Updating the Measurement Board (MPU)
1.
Download the new firmware image from the internet.
2.
Make sure that the "MPU RS232" lines of the serial interface connector are
connected with the serial interface of your computer.
3.
Start Teraterm.
4.
Switch the sensor on
on
on
on. Make sure the following message appears:
- Running self-test ...
- Selftest OK!
Type 4 x ENTER to switch to Terminal Mode
Check the serial connection if that message does not appear.
5.
Switch the sensor off.
6.
Choose Teratrem
File > Transfer > XModem > Send...
7.
7.
7.
7.
Locate and single click on the required firmware binary file.
8.
8.
8.
8.
Select 1K
1K
1K
1K from the options at the foot of the dialogue and click "
Open
""""::::