1.2
TRICOR Systems Inc.
1.2
Optional Items
The following are optional items that may be purchased for the Model 225:
•
Sample Cups (5000 / box)
•
External Printer with Serial Cable Adaptor (plotting option)
•
Temper Meter Data Acquisition Software (TMDAS)
•
USB Communications Cable
•
Calibration / Verification Fixtures
•
Optional Software Features
1.3
Serviceable Items
There are no user serviceable parts and no adjustments in the Model 225.
Opening the Model 225 voids the warranty and the Certificate of Calibration.
WARNING
There are dangerous voltages inside the Model 225
which may cause injury or death. Service of the Model
225 should be performed only by a qualified service
technician.
1.4
Calibration
TRICOR Systems recommends a yearly calibration. Two separate calibrations
are performed on Temper Meters. The first is an electronic calibration/
verification of the electronics. This verification can be performed by your internal
personnel provided that they have been given the proper training and the
verification tools have been purchased.
The second part of the calibration is a check of the heat transfer coefficients
between the cooler well and the sample cups. This calibration is performed at
TRICOR using specially crafted tools and procedures to obtain repeatable/
consistent results. This procedure takes 3 days provided that no repairs are
needed and the unit has been scheduled. An electronic calibration is performed
at the same time.