4.4
TRICOR Systems Inc.
4.2
Preferences (Cont'd)
The
Preferences
tab allows changes to the basic setup of the Model 225. This
includes the following features:
•
Line frequency
•
Password (to the menu)
•
Sample run time (option)
•
User text
•
Plot baseline temperature (option)
•
CTU offset
The
Features
group has a series of check boxes which indicate the availability
of various optional features. These are indicators only; they can not be clicked
on to change the state of the feature.
The
Allow User ID
check box indicates if the user is permitted to enter an ID
which is printed in the test results. The ID is a value of 1-199. The value is not
permanently stored in the Model 225 and can only be entered from the 225
menu. This ID can not be set from the Configuration application.
The
Multiple Run Storage
checkbox indicates the ability to store many
previous runs in the unit. The standard is to store only the last 5. The multiple
store uses a large memory to store to the capacity of the memory previous runs.
The number is a variable function that depends on the length of the run and the
number of aborted runs. Typical multiple storage results are:
Run Time
Capacity (Runs)
3:00
288
3:30
261
4:00
238
4:30
219
5:00 (standard)
202
5:30
188
6:00
176
6:30
165
7:00
156
7:30
147
8:00
140
8:30
133
9:00
127
9:30
121
9:55
117