3.20
TRICOR Systems Inc.
3.5.13 CTU Offset Adjust (Cont'd)
Press the
Select
button to select this menu group. Use the scroll buttons to
change the CTU Offset. The minimum value is -9.9 and the maximum is +9.9.
Holding the scroll buttons will make the value increment/decrement at a fast rate.
CTU Offset Adjust Menu
-1.0
To accept the newly entered value, press the
Select
button. If you decide not to
change the value, press the
Cancel
button to return without making the change.
In both cases the Top Menu Selection display will show again. To return to the
ready to test screen press the
Cancel
button.
3.5.14 Cooler Set Point
The Cooler Set Point Temperature is an option that must be purchased when
ordering the Model 225. If you do not have this option, selecting this menu group
will cause a warning message indicating that the feature is not available. This
menu group allows the cooler temperature to be adjusted.
Top Menu Selection
12 - Cooler Set Point
Press the
Select
button to select this menu group. Use the scroll buttons to
change the temperature. The temperature can be changed in 0.1 degree steps.
Cooler Set Point Menu
48.6
To accept the newly entered value, press the
Select
button. If you decide not to
change the value, press the
Cancel
button to return without making the change.
In both cases the Top Menu Selection display will show again. To return to the
ready to test screen press the
Cancel
button.