3.14
TRICOR Systems Inc.
3.5.7 Temperature Units (Cont'd)
To accept the newly entered value, press the
Select
button. If you decide not to
change the value, press the
Cancel
button to return without making the change.
In both cases the Top Menu Selection display will show again. To return to the
ready to test screen press the
Cancel
button.
NOTE
The CTU value has degrees as its unit of measure. Changing
the temperature units will affect the CTU values and the
stored CTU offset. Be sure that you correct the CTU offset
when switching between Celsius and Fahrenheit.
3.5.8 Set Clock/Calendar
The real time clock is an option that must be purchased when ordering the
Model 225. If you do not have this option, selecting this menu group will cause a
warning message indicating that the feature is not available. This menu group
allows the time and date for the Model 225 to be set.
NOTE
The Model 225 does not automatically correct for daylight
savings time. You must manually reset the time. The clock
is automatically corrected for leap years.
To change the time and date, press the
Select
button from the ready to test
screen. Use the scroll buttons to choose the Set Clock/Calendar group.
Top Menu Selection
6 - Set Clock/Calendar
Press the
Select
button to select this menu group. The current date and time will
be displayed. There will be a flashing cursor under the 2 digits that are to be
changed. Press the
Scroll Up
and
Scroll Down
buttons to change the date
and time.
Set Clock/Calendar Menu
09/27/04 10:27:00