FRONT
TOP
REMOVE THIS PANEL
FILL PORT
REAR VIEW
Chamber Fluid Filling Requires:
T10 Torx Driver
Funnel
3 U.S. gallons of propylene glycol
(4 U.S. gallons for 2500ST - supplied)
CHAMBER FLUID FILLING INSTRUCTIONS
Remove top right console panel (panel above chamber)
Locate fluid fill port (under insulation and labeled "Fluid Fill Port").
Remove red cap and insert funnel into fluid fill port.
Using 3 U.S. gallons of -100° C propylene glycol mix
(4 U.S. gallons for 2500ST) fill until level is approximately
1/4" below bottom of fill port, or to bottom of funnel.
Remove funnel and replace fluid fill port cap and insulation.
Replace top right console panel.
Chamber fluid filling is now complete.
1.
2.
3.
4.
5.
6.
7.
NO PROPYLENE
GLYCOL IN THIS PORT