2-1
Section 2
INSTALLATION
2.1 GENERAL
Preparations should be made to have adequate bench or floor space and a source of primary
power available at the location of use. A clean oil free instrument quality air supply must
also be available if the generator is installed without the cart air supply.
2.2 FACILITIES REQUIRED
Reference Drawing 91D25901
2.2.1 Bench / Floor Space
A bench and/or floor, capable of supporting approximately 300 pounds, with a minimum
space of 24" (610mm) deep x 40" (1.02m) wide is recommended and 27" (686mm) deep x
43" (1.09m) wide for 2500ST)(drawing 91D25901). Allow an additional 20" (508mm) in
width for clearance, if possible, so as to allow complete opening of the chamber door and
access to the test port.
Figure 2-1
2.2.2 Power
The 2500 humidity generator requires a primary power source of 120 VAC/60Hz or 100
VAC/50 Hz at 15 amps minimum (240 VAC/60 Hz or 220 VAC/50 Hz at 10 amps minimum
for HV option). The pneumatic cart with it's 1/2 horsepower air compressor requires an
additional power source of 120 VAC/60 Hz or 100 VAC/50 Hz at 8 amps minimum (240
VAC/60Hz or 220 VAC/50 Hz at 5 amps minimum for HV option).
COMPUTER / PRINTER
CHAMBER EXHAUST PORT
AIR SUPPLY INPUT (1/4" FPT)
MAIN POWER SWITCH (
CB
S1)
AC POWER
RESERVOIR FILL PORT
LEFT
RIGHT
FRONT
BACK
ACCESS PORT
INTERFACE
THUNDER SCI ENTIFIC COR POR ATI ON
2500 HUMIDI TY GENER ATOR