Benchtop Serial I/O
7/30/2015 3:30 PM
Page 11
Change Saturation Pressure Setpoint
PS=
Description
Changes the Saturation Pressure setpoint to a given value.
Prerequisites
Requires Benchtop firmware revision 1.05 or greater.
Syntax
PS=
setpoint
Parameters
setpoint
The
setpoint
value is the Saturation Pressure, in psia
only
, that the
system should control at.
Remarks
Control Mode Changed
Sending
PS=
setpoint
also changes the control mode to Saturation
Pressure (mode 3). In this mode, the %RH @Pc and %RH @PcTc
setpoints vary as system temperatures and chamber pressure change.
Data Returned
None. The only response is a carriage return/linefeed terminator.
Examples
The
PS=
setpoint
command is used to set the Saturation Pressure to
121.44 psia.
computer sends
PS=121.44
<CR>
benchtop responds
<CR><LF>
The
<CR><LF>
response signifies completion of the command.