2-5
2.3.2.4 Air Supply
Using a line size equal to or larger than 1/4" (6.35mm) OD, connect a source of clean, oil
free, instrument quality air (refer to 2.2.3) to the 1/4" FPT fitting located at the lower rear of
the console (working pressure is 175 PSIG). If the generator is installed on the pneumatic
cart, but an alternate air supply (facility) is desired, connect the alternate air supply (175 PSIG
MAWP) to the air inlet of the cart air supply so as to utilize the valves, storage tank, filter,
and 150 PSIG regulated output of the pneumatic cart (figure 2-5).
Figure 2-5
DV
LEFT SIDE
AIR INLET
BACK
A
T1
G1
BV
FRONT
REG
F2
F1
RIGHT SIDE