4-14
These new coefficients appear on the display.
13) To obtain a printout of the new coefficients, press [PRNT COEF].
14) To save the coefficients to non-volatile memory, press [SAVE QUIT]. To discard
these new coefficients and revert to the previous ones, press [EXIT QUIT].
15) At the calibration menu, press [DONE]. At the next menu, press [DONE] again. The
system reinitializes to the Control/Display screen.
16) To print a calibration report, refer to section 4.2.5.
4.2.5 Printing the Calibration Report
If a printer is connected to the Printer Port, a report may be printed which lists the calibration
coefficients and calibration dates for the temperature, pressure, and flow transducers. The
report is printed from the Edit/Cal menu.
To print the report:
1) From the Control/Display screen press [EDIT/CAL], or from the Cal menu press
[DONE]. Both actions should bring up the Edit/Cal menu.
2) Ensure the printer is on, then press [PRNT REPT]. The Calibration Report will be
sent to the printer.
3)
Press [DONE] to return to the Control/Display screen.
To save the report:
1) From the Control/Display screen press [EDIT/CAL], or from the Cal menu press
[DONE]. Both actions should bring up the Edit/Cal menu.
2) Using the computer cable supplied and a gender changer, connect a PC from its COM
port to the printer port of the 2500. Open the application “HyperTerminal” and
configure it to match the printer port parameters i.e. Baud, Data, Stop, Parity. Select
Capture Text from the Transfer menu and press [PRNT REPT] from the 2500 control
panel. When data output is finished, select Capture Text – Stop from the Transfer