4-8
7) Switch console power ON. Wait a few moments for the softkey menu to appear.
Allow approximately 30 minutes or more for warm-up of the pressure transducer
electronics.
8) Press the [EDIT/CAL] key, then the [CAL] key. The calibration menu appears.
9) Press the [PRES CAL] key.
10) Using [MARK/CLR] and the down arrow key as necessary, mark the transducer to be
calibrated. A marked transducer is indicated with an asterisk in the left most display
column. Since the two transducers require different calibration ranges, the computer
will only allow you to mark one transducer at a time.
11)
Press [OK], or if a mistake was made, press [EXIT QUIT]. [EXIT QUIT] reverts to
step 9.
TEMP
CAL
PRES
CAL
FLOW
CAL
DONE
Count PSIA
*_Low Range
High Range
MARK
/CLR
EXIT
QUIT
OK
Count PSIA
*Low Range
2445
12.18
High Range
818
12.19
Low
Mid
High
10
30
50
LOW
PRES
MID
PRES
HIGH
PRES
EXIT
QUIT