2-3
Figure 2-2
2.3.2 Positioning & Connections
Reference Drawings 91D25901, 09S25920, 91S25921, 98D25919 & 98D25925
2.3.2.1 Pneumatic Cart
Position the 2500 humidity generator on the cart (figure 2-3 & 2-4), with the cart push bar on
the left side of the generator, placing the leveling feet within the circular cutouts. Connect the
short air supply hose to the generator air supply inlet.
STANDARD COMPRESSOR
Remove the air compressor shipping strap. For operation, place the air compressor on
the floor and connect the 25 foot (7.62m) air supply hose between the air compressor
outlet and the cart air supply inlet.
Figure 2-3
FRONT
BACK
THUNDER SCI ENTIFIC COR POR ATI ON
2500 HUMIDI TY GENER ATOR
REMOVE THIS
PANEL
FILL PORT
REAR VIEW
FRONT
TOP
CHAMBER
SIDE
NO PROPYLENE GLYCOL
IN RESERVOIR FILL PORT!
.250" (6.35 mm)
PULL OUT
INSULATION
PLUG
FILL HERE
PULL OFF
RED CAP
CHAMBER
FILL PORT
CHAMBER
FILL FUNNEL
PROPYLENE GLYCOL
3 bottles for the R,
4 bottles for the ST