3.3 EDITING SYSTEM COEFFICIENTS AND PARAMETERS
All of the calibration coefficients and system parameters may be viewed and/or edited by the
operator. The following is a summary of items that may be edited.
Temperature Probe Calibration Coefficients
Pressure Transducer Calibration Coefficients
Flow Transducer Calibration Coefficients
Probe and Transducer Averaging
Serial Port Baud Rates, Parity, etc.
Print Interval and Lines Per Page
Time and Date
3.3.1 Edit Mode
The EDIT mode is used for the viewing and editing process.
1. From the Control/Display screen, press [EDIT/CAL]. Note, for this menu option to
appear, the generator must not be RUNNING. In a few seconds, the following menu
appears.
2. From the menu, press [EDIT].
3. At the prompt, enter in your authorization code (found at the back of the manual).