3-14
3.3.2.5 Console Port Parameters
These parameters affect the manner in which the bi-directional RS-232C Console Port
behaves. Changes become effective immediately.
Baud: 300, 600, 1200, 2400, 4800, 9600, 19200 or 38400 bits per second
Data: 7 or 8 bit word size.
Stop:
1 or 2 stop bits
Parity: 0 for NO parity, 1 for ODD parity, or 2 for EVEN parity
EOL: The ASCII value of the desired End-Of-Line or terminator character. Here, '13' is
the ASCII value for a Carriage Return.
Cancel: The ASCII value of the desired cancel-the-line character. Sending this character
clears anything in the input buffer of the Console Port. Here, '3' is the ASCII value
sent when executing a Control-C on most computers.
Caution:
It is advised that these settings stay at factory defaults.
The reason for keeping factory defaults is because the higher baud rates may cause overflow
and task overrun issues with the processor board.
Factory Default Settings:
Baud: 2400
Data: 8
Stop:
1
Parity: 0
EOL: 13
Cancel: 3
CONSOLE PORT PARAMETERS
Baud
[ 2400
]
Data
[ 8
]
Stop
[ 1
]
Parity
[ 0
]
EOL
[ 13
]
Cancel
[ 3
]
EDIT
NEXT
PREV
DONE