SOUND ENCLOSED COMPRESSOR (Optional)
Remove the sound enclosed air compressor shipping strap. For operation, the sound
enclosed air compressor may be left on the cart and connected to the cart air supply
inlet with the short air hose attached to it or may be set on the floor and connected
using the 25 foot (7.62m) air supply hose between the sound enclosed air compressor
outlet hose and the cart air supply inlet.
Figure 2-4
2.3.2.2 Bench Space
Position the 2500 humidity generator allowing at least 4 inches (102mm) of space near
console air intake (left side), and 4 inches (102mm) at the rear of the console for the exhaust
fan. Allow as much space as possible on the right side so as to allow clearance for utilizing
the chamber access port and to allow for opening of the chamber door. Level the console
using the top as a reference. Tighten leveling leg locking nuts against frame.
2.3.2.3 Power
Be sure the main console Power switch is OFF, then plug the 2500 into a source of 120
VAC/60Hz or 100 VAC/50 Hz at 15 amps minimum (240 VAC/60 Hz or 220 VAC/50 Hz at
10 amps minimum for HV option). If using the pneumatic cart, plug the air compressor into a
source of switched 120 VAC/60Hz or 100 VAC/50 Hz at 8 amps minimum (240 VAC/60 Hz
or 220 VAC/50 Hz at 5 amps minimum for HV option).
G1
AT1
BV1
FRONT
RIGHT SIDE
REG
F2
F1
BV2
BACK
DV
LEFT SIDE
ACS
AH1
AH2
2-4