Maintaining Your Instrument
Instrument Diagnostics
Thermo Scientific
NanoDrop One User Guide
293
To interpret intensity check results
If one of these indicators:
• UV
• Visible
• Bias
has an adjacent yellow triangle instead of the green check marks shown above,
pedestals with deionized water
and then repeat the Intensity Check.
If a yellow triangle appears next to the Bias indicator, make sure the room is within the
temperature specifications for the instrument.
If the Intensity Check fails again,
Related Topics
Performance Verification
Run Performance Verification every 6 months to confirm pathlength accuracy is within
specifications.
Supplies needed
• lint-free laboratory wipes
• deionized water (DI H
2
O)
• calibrated precision pipettor (0–2 μL)
•
PV-1 performance verification solution
(liquid photometric standard available only from
us or a local distributor)
• laboratory gloves
Note
The PV-1 solution comes in a single-use ampoule. Before you open the ampoule,
shake it vigorously and then allow the liquid to collect in the bottom portion of the
ampoule. After the ampoule is opened, its contents must be used within one hour. Pipette
directly from the ampoule; do not transfer the solution.
Summary of Contents for NanoDrop One
Page 18: ......
Page 28: ......
Page 58: ......
Page 134: ...Measure Protein Pierce 660 128 NanoDrop One User Guide Thermo Scientific Instrument Settings ...
Page 144: ......
Page 152: ......
Page 159: ...Measure UV Vis Thermo Scientific NanoDrop One User Guide 153 ...
Page 160: ...Measure UV Vis 154 NanoDrop One User Guide Thermo Scientific ...
Page 239: ...3 Learning Center Instrument Settings Thermo Scientific NanoDrop One User Guide 239 ...
Page 312: ...7 Contact Technical Support 314 NanoDrop One User Guide Thermo Scientific ...