Protein Pierce 660 Reported Results
Protein Pierce 660 measurement screen (shown from Data
Viewer)
For each measured sample and standard, this application shows the visible absorbance
spectrum and a summary of the results. The standard curve is also available by swiping left
from the measurement screen (or in the
as shown below).
Tap row
to
select sample
and update
spectrum;
tap
more rows
to
overlay up to five
spectra.
Press
and hold
sample row to
view
measurement
details.
Drag tab down/up
to see
more/less sample data
Swipe screen left
to
view standard curve
Pinch and zoom
to adjust axes;
double-tap to reset
Visible
spectrum
Menu of options;
tap
to open
Protein
concentration
Absorbance
at 660 nm
Note
• A baseline correction is performed at 750 nm (absorbance value at 750 nm is
subtracted from absorbance values at all wavelengths in sample spectrum).
• Micro-volume absorbance measurements and measurements taken with nonstandard
cuvettes are normalized to a 10.0 mm pathlength equivalent.
Summary of Contents for NanoDrop One
Page 18: ......
Page 28: ......
Page 58: ......
Page 134: ...Measure Protein Pierce 660 128 NanoDrop One User Guide Thermo Scientific Instrument Settings ...
Page 144: ......
Page 152: ......
Page 159: ...Measure UV Vis Thermo Scientific NanoDrop One User Guide 153 ...
Page 160: ...Measure UV Vis 154 NanoDrop One User Guide Thermo Scientific ...
Page 239: ...3 Learning Center Instrument Settings Thermo Scientific NanoDrop One User Guide 239 ...
Page 312: ...7 Contact Technical Support 314 NanoDrop One User Guide Thermo Scientific ...