Measure using a Custom Factor
40
NanoDrop One User Guide
Thermo Scientific
For example, for a sample measurement using an extinction coefficient of 55, the equation
looks like this:
(550 AU * 55 ng-cm/μL) = 30,250 ng/μL
Related Topics
• Detection Limits for All Applications
Note
For measurements with 10 mm pathlength cuvettes, the upper absorbance limit is
1.5 AU, which is approximately 75 ng/μL for dsDNA.