Measure Kinetics
Thermo Scientific
NanoDrop One User Guide
161
Kinetics Reported Results
Absorbance measurement screen
The absorbance measurement screen appears immediately after you tap Measure in the
kinetics experiment. This screen shows the absorbance spectrum for each measurement, with
wavelength on the X-axis and absorbance on the Y-axis. Vertical lines indicate the specified
wavelengths to monitor. The list at the right shows the time each measurement was taken in
each specified stage (drag the tab down to see more entries). Each item in the list at the right
has a corresponding absorbance spectrum at the left. The image below highlights the available
features.
Tap row
to
select
measurement
and update
spectrum;
tap
more rows
to
overlay up to five
spectra.
Press
and hold
row to
view
measurement
details.
Drag tab
down/up
to see
more/less
sample data
Swipe screen left
to view
Rate measurement screen
Pinch and zoom
to adjust
axes; double-tap to reset
Tap
to end experiment
and export data
Menu of options;
tap
to open
Method name
User-defined wavelengths
(i.e., 260 nm, 340 nm, 660 nm)
Note
For measurements taken with nonstandard (other than 10 mm) cuvettes, the spectra
are normalized to a 10 mm pathlength equivalent.
Summary of Contents for NanoDrop One
Page 18: ......
Page 28: ......
Page 58: ......
Page 134: ...Measure Protein Pierce 660 128 NanoDrop One User Guide Thermo Scientific Instrument Settings ...
Page 144: ......
Page 152: ......
Page 159: ...Measure UV Vis Thermo Scientific NanoDrop One User Guide 153 ...
Page 160: ...Measure UV Vis 154 NanoDrop One User Guide Thermo Scientific ...
Page 239: ...3 Learning Center Instrument Settings Thermo Scientific NanoDrop One User Guide 239 ...
Page 312: ...7 Contact Technical Support 314 NanoDrop One User Guide Thermo Scientific ...