NanoDrop One Viewer Software
Thermo Scientific
NanoDrop One User Guide
269
Custom method settings
These settings are available for creating custom methods.
Setting
Available Options
Measurement range
Select spectral range in which method will acquire data.
Available options:
• Ultra-violet only (190 nm - 350 nm)
• Visible only (350 nm - 850 nm)
• Ultra-violet and visible (190 nm - 850 nm)
• Custom (specify starting and ending point in nanometers)
Notes
:
• If a
and/or
Analysis wavelength correction
your selected spectral range includes your specified baseline correction and/or
analysis correction wavelength.
• For micro-volume absorbance measurements and measurements taken with
nonstandard (other than 10 mm) cuvettes, the spectra are normalized to a 10 mm
pathlength equivalent.
Analysis wavelength
Monitor absorbance at specified wavelength (enter the wavelength in nanometers).
Note
: The specified wavelength must fall within the selected
The measurement results or the concentration will be calculated automatically using
the absorbance value at the specified wavelength and applying the selected method
type (factor or standard curve).
Result name
Enter descriptive name for calculated concentration result (for example, “Nucleic
Acid”) and use adjacent drop down list to select appropriate unit. Result name appears
as column heading for reported concentration value.
Summary of Contents for NanoDrop One
Page 18: ......
Page 28: ......
Page 58: ......
Page 134: ...Measure Protein Pierce 660 128 NanoDrop One User Guide Thermo Scientific Instrument Settings ...
Page 144: ......
Page 152: ......
Page 159: ...Measure UV Vis Thermo Scientific NanoDrop One User Guide 153 ...
Page 160: ...Measure UV Vis 154 NanoDrop One User Guide Thermo Scientific ...
Page 239: ...3 Learning Center Instrument Settings Thermo Scientific NanoDrop One User Guide 239 ...
Page 312: ...7 Contact Technical Support 314 NanoDrop One User Guide Thermo Scientific ...