Unpacking and Installation
Thermo Fisher Scientific
SAM12 User Manual 4-5
“bedding-in” adjustments as described in
Door Locking
Mechanism (if fitted) (page 8-1)
.
Electrical Installation
Before proceeding with the electrical installation, ensure the
SAM12 has been mechanically installed as described in
Mechanical Installation (page 4-2)
.
IMPORTANT:
ENSURE CORRECT ELECTRICAL SUPPLY
AND MAINS CONNECTOR WIRING FOR
YOUR MAINS SUPPLY.
SAM12 is shipped with the lead-acid battery discharged.
The SAM12 is fitted with a 50 VA auto ranging power supply
requiring a maximum input power of 65 VA.
DO NOT CONNECT THE MAINS SUPPLY. ENSURE
THE KEYSWITCH ON THE REAR PANEL IS SET TO
THE OFF POSITION.
A standard IEC mains connector is fitted on the side panel,
specified to IEC 320, EN60320. Only mating connectors of the
SAM12 standard should be used.
The SAM12 must be connected to a fused outlet capable of
being isolated. The isolation switch should be close to the
monitor and within reach of the operator.
The IEC power cord must be connected to the electricity supply
as follows:
Brown conductor
to
LIVE ‘L’
or
Red terminal
Blue conductor
to
NEUTRAL ‘N’
or
Black terminal
Green/Yellow
conductor
to
EARTH ‘E’
GROUND
or
Green/Yellow
terminal
WARNING: THIS EQUIPMENT MUST AT NO TIME BE
OPERATED WITHOUT THE GREEN WIRE CONNECTED
TO THE MAINS GROUND.
When all installation work has been carried out, follow the
setting up procedure described in
Setting Up Procedure (page
.
Battery State
Electrical Installation
Summary of Contents for SAM12
Page 1: ...MANUAL SAM12 SMALL ARTICLES MONITOR Thermo Fisher Scientific SAM12 21st September 2007...
Page 2: ......
Page 4: ...This manual was produced using ComponentOne Doc To Help...
Page 10: ......
Page 18: ......
Page 20: ......
Page 34: ......
Page 74: ...Operating Instructions Issue 1 0 5 34 SAM12 User Manual Thermo Fisher Scientific...
Page 130: ......
Page 134: ......
Page 144: ......
Page 168: ......