Operating Instructions
Issue: 1.0
5-26 SAM12 User Manual
Thermo Fisher Scientific
amplifiers at each HV step over the specified integration
period:
•
A background HV Scan
•
A source HV Scan
When the options have been
entered, press Continue button to
display Ready to Start
Background Scan window:
Remove any current Source from
the SAM12, close the doors and
press the Start button. A
background HV scan will be
performed, recording the counts
against
each amplifier threshold
for all amplifiers at each HV step
over the specified integration
period. A progress window is
displayed:
Once the Background Scan is
completed, the Ready to Start
Scan of Source is displayed:
Insert the Source to be scanned
into the SAM12, close the doors
and press the Start button. A
source HV scan will be
performed, recording the counts
against
each amplifier threshold
for all amplifiers at each HV step
over the specified integration
period. A progress window is
displayed:
Once the Scan is completed, a Save Scan button appears giving
the option to save the scan or Abort the process.