Operating Instructions
Issue: 1.0
5-10 SAM12 User Manual
Thermo Fisher Scientific
message displayed on the unit when out of date and
for information of the last
Calibration Date.
Values: 1 to 400 days (0 to disable)
Default: 0
Max Clear-Result
Display Time (s)
This is the maximum length of time (in seconds)
that the Clear result is displayed before reverting to
the Ready screen.
Values: 0 to 500 s
Default: 300 s
For further information regarding these options, contact
Thermo Fisher Scientific.
Select Set-Up|Params2 to
customise the second set of
Operational Parameters
Description of options:
Field Name
Description
Probability of
Detection (
σ
)
This is the probability that exactly one alarm Level
of contamination will cause and alarm.
For further information, see
Description of
Parameters Used in Calculations (page 6-2)
.
Note that the probability is also expressed as a
percentage in parentheses
Values: 0 to 10 sigma in 0.1 sigma steps.
Default: 1.65
σ
Probability of False
Alarm (
σ
)
This value is the probability that a false alarm will
NOT be given during a measurement. This
probability is used by SAM12 to determine the
monitoring time and the Effective Alarm Level and
Minimum Detectable Activity.
For further information, see
Description of
Parameters Used in Calculations (page 6-2)
.
Note that for simplicity, the Probability of False
Alarm is expressed as a percentage in parentheses.
Values: 0 to 10 sigma in 0.1 sigma steps.
Default: 3.1
σ
Minimum Monitoring
Time (s)
Note that the Minimum monitoring time must be
equal to or be less than Maximum Monitoring Time.
This is the Minimum time for which articles are
monitored. This field is used in conjunction with
the Maximum Monitoring Time.
Values: 3 to 300 s
Params 2
Summary of Contents for SAM12
Page 1: ...MANUAL SAM12 SMALL ARTICLES MONITOR Thermo Fisher Scientific SAM12 21st September 2007...
Page 2: ......
Page 4: ...This manual was produced using ComponentOne Doc To Help...
Page 10: ......
Page 18: ......
Page 20: ......
Page 34: ......
Page 74: ...Operating Instructions Issue 1 0 5 34 SAM12 User Manual Thermo Fisher Scientific...
Page 130: ......
Page 134: ......
Page 144: ......
Page 168: ......