Operating Instructions
Issue: 1.0
5-56 SAM12 User Manual
Thermo Fisher Scientific
Note that if the number of counts
is less than Minimum Detectable
Activity – see
- the Clear result will state
that it is less than the MDA.
After a CLEAR result, the user is prompted to remove the
measured article. When the door is opened to retrieve the
article, the "READY – Deposit Article and Close Door"
message is displayed – see
to start
a new measurement.
CAUTION: THE SAM12 CANNOT DISTINGUISH
BETWEEN A GRADUAL BUILD-UP OF
CONTAMINATION AND SMALL BACKGROUND
CHANGES. THEREFORE, IT IS STRONGLY
RECOMMENDED THAT A REGULAR FRISKING
PROCEDURE IS ADOPTED AND/OR AN INSPECTION
OF THE BACKGROUND REPORT SHOULD INDICATE
A BUILD-UP OF ACTIVITY – SEE
.
If the measurement completes but
the result is ALARM or when
QuickScan indicates real-dirty,
the user is informed audibly and
visually of the result.
Note that this mess
age is user-
definable. See
for mor
e information.
For more information, see
.
Note that the Uncertainty Value
and Activity results are displayed
if enabled – see
for more information.
ALARM Result