Operating Instructions
Issue: 1.0
5-30 SAM12 User Manual
Thermo Fisher Scientific
Note that this menu option is only displayed if the CCM
Enabled option is ticked in the Options page - see
(page 5-6)
.
Select CCM|Setup to select the
parameters for the CCM Channel
Note that this section is set by at production and should only be
changed in consultation with Thermo Fisher Scientific.
Description of options:
Field Name
Description
Coincidence
Threshold (mV)
This is the high threshold for summed pulses.
Coincidence Period
(ns)
This is an estimate of the Coincidence Period set in
nanoseconds.
Select CCM|Information to view
the status information for the
CCM Channel
Description of options:
Field Name
Description
MDA (default units)
Note that this is a view only field.
The Minimum Detectable Activity (MDA) is a
value derived by the SAM12 and is the smallest
activity that can be statistically accurately detected
with the current settings and background. The
MDA is calculated using the probability of
detection, probability of false alarm, monitoring
time and the average background etc. Dashes are
shown if this value is not available.
Detection Limit (cps)
Note that this is a view only field.
This is the MDA expressed in counts per second
CCM Menu
CCM Setup
Information
Summary of Contents for SAM12
Page 1: ...MANUAL SAM12 SMALL ARTICLES MONITOR Thermo Fisher Scientific SAM12 21st September 2007...
Page 2: ......
Page 4: ...This manual was produced using ComponentOne Doc To Help...
Page 10: ......
Page 18: ......
Page 20: ......
Page 34: ......
Page 74: ...Operating Instructions Issue 1 0 5 34 SAM12 User Manual Thermo Fisher Scientific...
Page 130: ......
Page 134: ......
Page 144: ......
Page 168: ......