Calibration Procedure
Issue: 1.0
10-2 SAM12 User Manual
Thermo Fisher Scientific
Since the SAM has a very linear response to activity,
any source of activity between 100 nCi (3.7 kBq) to 10
μ
Ci (370 kBq) range may be used. When CCM is
enabled, the 60Co source used for calibration should
have an activity in excess of 5
μ
Ci (185 kBq) in order
to minimise the calibration time. Long count times will
be required for small sources to maintain statistical
accuracy against adverse background influences.
2.
A source holder, to securely retain the source for
positioning within the cubicle.
3.
A generic USB printer, (e.g. Epson LX300)
Preparation for Calibration
First the user must enter the
Calibration Menu (page 5-19)
.
Select
and check the detector HV’s are
set correctly for the Calibrating isotope. If these are not
known, they must be determined as described in
Detector HV
Selection (page 9-3)
before proceeding with the calibration.
The SAM12 is then ready for calibration.
Calibration
Ensure the background is stable, the door is closed and any
calibration sources are well away from the SAM12 (at least 8
feet).
1.
Ensure the door is shut and the unit is on and ready.
2.
Login to the Administration mode, using the Health
Physicist role (see
Gaining Access to the
Administration Mode (page 5-3)
).
See
for more information.
To check the calibration, the
Cal Check
button is selected. This
does not allow the calibration factor to be changed, but a
comparison is made between the old and new factor.
To recalibrate the instrument, i.e. adjust the calibration factor
for a particular nuclide, the
Calibrate
button is selected.
The calibration process should be repeated for a number of
nuclides that would be found in typical contamination. As a
minimum, a calibration to both
137
Cs and
60
Co should be
undertaken. The
137
Cs calibration is essential when the NBR
facility is used; the
60
Co calibration is essential when CCM is
required.
Notes: Ensure that each source is at least 8 feet away from the
SAM during the background measurement.
Summary of Contents for SAM12
Page 1: ...MANUAL SAM12 SMALL ARTICLES MONITOR Thermo Fisher Scientific SAM12 21st September 2007...
Page 2: ......
Page 4: ...This manual was produced using ComponentOne Doc To Help...
Page 10: ......
Page 18: ......
Page 20: ......
Page 34: ......
Page 74: ...Operating Instructions Issue 1 0 5 34 SAM12 User Manual Thermo Fisher Scientific...
Page 130: ......
Page 134: ......
Page 144: ......
Page 168: ......