Issue: 1.0
Operating Instructions
Thermo Fisher Scientific
SAM12 User Manual 5-31
(for the CCM Channel only).
Measurement Period
(s)
Note that this is a view only field.
This is the length of time for the sample calculated
from the current alarm conditions and background
measurement (for the CCM Channel only).
Background Rate
(cps)
Note that this is a view only field.
This is the current background rate (for the CCM
Channel only). The value is set at 0 if a background
has not yet been acquired.
CCM Efficiency (%)
Note that this is a view only field.
This is the Efficiency for the CCM Channel.
Net Effective Alarm
Rate (cps)
Note that this is a view only field.
This is the effective alarm rate for the CCM
Channel calculated using the CCM Alarm setting
taking account of current alarm conditions and
background measurement (see
).
Select CCM|Diagnostics to view
the activity on the CCM Channel.
This function allows a count to
be made which can be used to
ascertain the efficiency and
functionality of the SAM12:
To Start the Timed Counts for the Detectors
:
1.
Select the number of SAM12 to be averaged by
selecting a value in the Average Over field.
2.
If you require a Continuous count to be made, tick the
Continuous box.
3.
Press Start button
Note that the Arbitrary field displays the pulses obtained by
random coincidence from non-Cobalt60 source.
Select Data|Reports to select a
Database Report from a pre-
defined list:
Diagnostics
Data Menu
Reports
Summary of Contents for SAM12
Page 1: ...MANUAL SAM12 SMALL ARTICLES MONITOR Thermo Fisher Scientific SAM12 21st September 2007...
Page 2: ......
Page 4: ...This manual was produced using ComponentOne Doc To Help...
Page 10: ......
Page 18: ......
Page 20: ......
Page 34: ......
Page 74: ...Operating Instructions Issue 1 0 5 34 SAM12 User Manual Thermo Fisher Scientific...
Page 130: ......
Page 134: ......
Page 144: ......
Page 168: ......