Maintenance and Trouble Shooting
Issue: 1.0
11-18 SAM12 User Manual
Thermo Fisher Scientific
1.
Check the cable connections between the Touch screen
keypad and the Controller board.
2.
Temporarily plug a spare LCD assembly into the
control board. If this solves the problem the original
Touch-pad is faulty and should be replaced as described
in
LCD Display, Touch Screen, Touch Controller &
Backlight Inverter - Removal and Replacement (page
11-8)
3.
The fault is likely to be in the Controller board which
should be replaced as described in
.
If, after power-up, or after exiting from the main menu, the
SAM12 displays “Cannot Measure Background - Please Close
the Door”, a fault with the door switch system is likely.
Proceed as follows:
1.
Check that the door switch is working correctly by
selecting “Diagnostics/Vault”. The switch is situated in
the left hand vertical box section of the main frame and
is operated by closing the door. Replace any switches if
necessary.
2.
Check the continuity of the cable connecting the door
switch(s) to the Controller board.
Table 4 - Summary of the Controller Board Inputs
Input Connection Function
IP1 PL24
Door
Switch
(dirty)
IP2
PL27
Solenoid Status (dirty)
IP3 PL28
Door
Switch
(clean)
IP4
PL29
Solenoid Status (clean)
IP5 PL20
Start
Button
Replace the Controller board as described in
type 5670A - Removal and Replacement (page 11-6)
if the
switches are working correctly.
If the SAM12 has acquired a valid set of background readings,
but will not enter measurement mode when a sample is entered
and the Start switch is operated, proceed as follows:
1.
Check that the Start switch is working correctly by
using “Diagnostics/Vault”. The switch is situated in the
left hand side of the front panel. Replace if necessary.
Will not Enter Background
Mode
Locked in Background
Mode
Summary of Contents for SAM12
Page 1: ...MANUAL SAM12 SMALL ARTICLES MONITOR Thermo Fisher Scientific SAM12 21st September 2007...
Page 2: ......
Page 4: ...This manual was produced using ComponentOne Doc To Help...
Page 10: ......
Page 18: ......
Page 20: ......
Page 34: ......
Page 74: ...Operating Instructions Issue 1 0 5 34 SAM12 User Manual Thermo Fisher Scientific...
Page 130: ......
Page 134: ......
Page 144: ......
Page 168: ......