Issue: 1.0
Operating Instructions
Thermo Fisher Scientific
SAM12 User Manual 5-13
Default: 5000.0 dpm
High Level Alarm
(default unit)
This is the detected activity at which SAM12 will
indicate an High Level Alarm
The Enabled tickbox must be checked for this
option to take effect.
Values: 0 to 10,000,000 (Any unit).
Default: Must be input
CCM Alarm (default
unit)
This field is enabled if CCM Enabled is ticked - see
.
This is the detected activity at which SAM12 will
indicate a CCM Alarm.
Values: 0 to 10,000,000 (Any unit).
Default: Must be input
CCM High Level
Alarm (default unit)
This field is enabled if CCM Enabled is ticked - see
.
This is the detected activity at which SAM12 will
indicate a CCM High Level Alarm.
The Enabled tickbox must be checked for this
option to take effect.
Values: 0 to 10,000,000 (Any unit).
Default: Must be input
Select Set-Up|Doors to select the
Number of Doors, whether Locks
are fitted and their behaviour
Note that this section is set by at production and should only be
changed in consultation with Thermo Fisher Scientific.
Description of options:
Field Name
Description
Number of Doors
Use the Radio buttons to select either One Door or
Two Doors.
Default: Two
Locks Fitted :
Front Door
Tick this box if the Front Door has a lock fitted.
Also referred to as the "Dirty" door.
Default: ON
Rear Door
Note that this option is only visible if Two Doors is
selected in Number of Doors.
Tick this box if the Rear Door has a lock fitted.
Also referred to as the "Clean" door
Doors
Summary of Contents for SAM12
Page 1: ...MANUAL SAM12 SMALL ARTICLES MONITOR Thermo Fisher Scientific SAM12 21st September 2007...
Page 2: ......
Page 4: ...This manual was produced using ComponentOne Doc To Help...
Page 10: ......
Page 18: ......
Page 20: ......
Page 34: ......
Page 74: ...Operating Instructions Issue 1 0 5 34 SAM12 User Manual Thermo Fisher Scientific...
Page 130: ......
Page 134: ......
Page 144: ......
Page 168: ......