Operating Instructions
Issue: 1.0
5-52 SAM12 User Manual
Thermo Fisher Scientific
RESTRAINT; IT MAY BE DANGEROUS AND CAUSE
DAMAGE TO THE SAM12.
CAUTION: THE DOORS ARE VERY HEAVY AND
MUST BE CLOSED WITH REASONABLE CARE, FOR
SAFETY REASONS AND TO AVOID UNIT DAMAGE.
See
for more information on
manually setting up the doors.
With the instrument in the Ready
state, displaying "READY –
DEPOSIT ARTICLE AND
CLOSE DOOR", the door can be
opened to commence the
monitoring of an article (Note
that "Door Open" is displayed in
the Operational Mode display
bar).
Note that when the door is
opened, the background
monitoring is suspended.
Deposit the article to be
monitored in the SAM12 and
close the door. (Note that "Door
Closed Waiting for Start" is
displayed in the Operational
Mode display bar).
If the door is not closed within 10
seconds, a repeating alarm shall
be sounded.
The "Ready – PUSH START
BUTTON TO BEGIN
MONITORING" message is
displayed.
The “ILLEGAL DOOR
SEQUENCE – Please close both
doors” message is displayed
under the following
circumstances:
- If the back door is opened on a
two-door SAM12 when
attempting to deposit an article.
- If both doors are opened at the
same time on a two-door
SAM12.
Notes regarding Illegal Door
Operations
Summary of Contents for SAM12
Page 1: ...MANUAL SAM12 SMALL ARTICLES MONITOR Thermo Fisher Scientific SAM12 21st September 2007...
Page 2: ......
Page 4: ...This manual was produced using ComponentOne Doc To Help...
Page 10: ......
Page 18: ......
Page 20: ......
Page 34: ......
Page 74: ...Operating Instructions Issue 1 0 5 34 SAM12 User Manual Thermo Fisher Scientific...
Page 130: ......
Page 134: ......
Page 144: ......
Page 168: ......